Device Dossier
Senza Omnia IPG Kit - Analgesic spinal cord electrical stimulation system pulse generator implantable
Device Dossier for Senza Omnia IPG Kit - Analgesic spinal cord electrical stimulation system pulse generator implantable (Analgesic spinal cord…
Intended purpose
The Senza neuromodulation systems are intended to aid in the management of chronic intractable pain of the trunk and/or limbs, including unilateral or bilateral pain associated with the following: failed back surgery syndrome, intractable low back pain, upper back pain and leg pain.
Functional description
The IPG is a rechargeable implantable device with 16 output channels capable of stimulating the spinal cord nerves through electrode leads. The IPG is designed to produce current-regulated, charge-balanced, biphasic, capacitively-coupled, rectangular output pulses. The IPG Header contains the charging coil and two ports to allow the insertion of leads. The rechargeable battery is contained inside the IPG titanium enclosure. The kit contains IPG, torque wrench, and two port plugs.